Starters for cage motors
  Starters for slipring motors
 
 
 
 
Gas
 

 

N°2.1525 * Temperature

N°2.1524 Electricity -magnetism

*Range on www.cofrac.fr

 
 
Home > Instrumentation > Calibration Measurement > Field instruments >

Back
DOCUMENTING PROCESS MULTIFUNCTION SIMULATORS: CALYS IS 60 / 80 / 120 for ATEX Area  
   
    ATEX Ex II 1G EEx ia IIC T4
 
Protected for field use    
Memory    
 

For accurate field and lab measurement
Use CALYS documenting process simulator -900g only & large dual display- to accurately perform, with one single instrument, all field and lab measurements:
  • Temperature: Up to 0.01 % RDG
  • Resistance: Up to 0.008 % RDG and 5 KΩ range
  • Current: Up to 0.01 % RDG and 100 mA range + Loop Supply 12 V
  • Voltage: Up to 0.006 % RDG and 50 V range
  • Frequency: Up to 0.005 % RDG and 100 KHz range
  • Pressure: Internal upto 20 bars and extrenal up to 700 bars. Accuracy up to 0.025 % FS. Relative and absolute pressure ranges
 
Advanced features

Dual display IN / OUT:Enable dual reading of the two channels values

Memory: Store your calibration procedures in the device and save your results for on site easy and quick calibration during the whole week. Then, upload the data on a computer in order to issue calibration certificates

Additional functions: Scaling, steps, synthetizer, statistical functions... make advanced data exploitation easier

CALYS
 
Easy connection system
Easy Connect Push on the terminal top and insert wires with a diameter of up to 3 mm or 10 AWG and compensated thermocouple connectors.

Wires are held tight between two brass plates that provide thermal stability and a very good cold junction compensation for thermocouples.

This system also enables 4 mm banana plugs and security connectors to be connected on the front panel.
  Request For Quote
  PDF CALYS IS data sheet
 
CALYS
Size: 290 x 98 x 507 mm
Waterproof: IP54
Weight: 1,4 kg
Operating conditions: 0 / 50° C, 10 / 80% RH
3 models from basic use to advanced technology
Specifications CALYS 60 CALYS 80 CALYS 120
Top accuracy 200 ppm 100 ppm 60 ppm
Temperature accuracy Thermocouples (14)
RTDs (10)
0.02% RDG for Tc K
0.02% RDG for RTD
0.01% RDG for Tc K
0.01% RDG for RTD
0.01% RDG for Tc K
0.01% RDG for RTD
DC current
+ Loop supply 12 V
Range
Accuracy
Up to 50 mA
0.02% RDG
Up to 50 mA
0.01% RDG
Up to 50 mA
0.01% RDG
DC voltage Range
Accuracy
12 V IN / OUT
0.02% RDG
12 V IN / OUT
0.01% RDG
12 V IN /: OUT
0.006% RDG
Frequency Range
Accuracy
Up to 20KHz
0.005% RDG
Up to 20KHz
0.005% RDG
Up to 20KHz
0.005% RDG
Resistance Range
Accuracy
Up to 5000 Ω
0.02% RDG
Up to 5000 Ω
0.01% RDG
Up to 4000 Ω
0.008% RDG
Pressure Range
Accuracy
 

Relative pressure: Up to 2 bar(Internal)/700 bar (External)

Absolute pressure: Up to 20 bar (External)
0.025% RDG

Additional functions Advanced data exploitation: Scaling, relative measurement, simulation of ramps and steps, square root, Calibration of transmitters
  Transmitter function, switch test
Memory
    Dual Input channels for calibration by comparison Method
Communication   USB connexion
Power supply Rechargeable NiMh batteries
 
 
 
Home |Company | Events | Support |Contact  | © All rights reserved 2008