| Specifications |
CALYS 60 |
CALYS 80 |
CALYS 120 |
| Top accuracy |
200 ppm |
100 ppm |
60 ppm |
| Temperature accuracy |
Thermocouples (14)
RTDs (10) |
0.02% RDG for Tc K
0.02% RDG for RTD |
0.01% RDG for Tc K
0.01% RDG for RTD |
0.01% RDG for Tc K
0.01% RDG for RTD |
DC current
+ Loop supply 12 V |
Range
Accuracy |
Up to 50 mA
0.02% RDG |
Up to 50 mA
0.01% RDG |
Up to 50 mA
0.01% RDG |
| DC voltage |
Range
Accuracy |
12 V IN / OUT
0.02% RDG |
12 V IN / OUT
0.01% RDG |
12 V IN /: OUT
0.006% RDG |
| Frequency |
Range
Accuracy |
Up to 20KHz
0.005% RDG |
Up to 20KHz
0.005% RDG |
Up to 20KHz
0.005% RDG |
| Resistance |
Range
Accuracy |
Up to 5000 Ω
0.02% RDG |
Up to 5000 Ω
0.01% RDG |
Up to 4000 Ω
0.008% RDG |
| Pressure |
Range
Accuracy |
|
Relative pressure: Up to 2 bar(Internal)/700 bar (External)
Absolute pressure: Up to 20 bar (External)
0.025% RDG |
|
|
| Additional functions |
Advanced data exploitation: Scaling, relative measurement, simulation of ramps and steps, square root,
Calibration of transmitters |
| |
Transmitter function, switch test
Memory |
| |
|
Dual Input channels for calibration by comparison Method |
|
| Communication |
|
USB connexion |
| Power supply |
Rechargeable NiMh batteries |