| Functionalities |
OM 16 |
| Measurement: |
Low resistance value measurement in 4-wire configuration
Inductive resistances
Non-inductive resistances |
| Ranges: |
7 ranges from 5m to 2.5 KΩ, manual and auto ranging |
| High accuracy: |
0.05 % RDG |
| Current: |
DC current, from 1mA to 10 A |
| High resolution: |
0.1 µΩ |
| Low temperature coefficient: |
10 ppm/°C |
| Operating modes: |
Manual or automatic measurement trigger:
one single operator can perform the measurement |
| Resistance calculation: |
EMF measured before each measurement Display of temperature compensated resistance value Choice of metal temperature coefficient Choice of reference temperature and ambient temperature (programmed or measured with integrated sensor) |
|
| Power supply: |
Mains and quick charge NiMh batteries |
| Communication modes: |
RS 232 |
| Protection: |
Up to 250 V |
|
| Additional functions: |
Two programmable thresholds
Digital calibration without internal adjustment
Storage capacity: 1000 values identified to read on display and via Log OM software or printer |
| Applications: |
Inductive resistances: coils, transformers, motor windings...
Non-inductive resistances: earth bonding, coating, contact resistances... |