Starters for cage motors
  Starters for slipring motors
 
 
 
 
Gas
 

 

N°2.1525 * Temperature

N°2.1524 Electricity -magnetism

*Range on www.cofrac.fr

 
 
Home > Instrumentation > Calibration Measurement > Field instruments >

Back
DUAL INPUT THERMOMETER / CALIBRATOR: THERMYS150 + Calibration management software: DATACAL
21 CFR Part 11
0.005 % RDG accuracy in measurement and simulation of temperature NADCAP Heat treatment compliant
Comparison calibration
Differential thermal analysis
 

THERMYS 150 is a all-in-one thermometer / calibrator for thermocouples, resistive probes and thermistors providing a high accuracy:
  • Thermocouples: Up to 0.005 % RDG
  • RTDs: Up to 0.006 % RDG
  • Thermistors: Up to 0.006 % RDG
Two input channels are available on the instrument for comparison calibration and temperature differential measurement.. Temperature can also be simulated by the output channel in order to control temperature recorders, indicators, handheld thermometers...
 
Advanced features
  • Memory: Store your calibration procedures in the device and save your results for on site easy and quick calibration during the whole week. Then, upload the data on a computer in order to issue calibration certificates
  • Customized calibration certificates: Issue your own customized calibration certificates with DATACAL calibration software (see below)
  • Additional functions: Scaling, steps, synthetizer, statistical functions, switch tests... make advanced data exploitation easier
  • Driving dry-blocks and baths: Furnaces can be driven by THERMYS150 through a 4-20 mA signal sent by the output channel in order to set up the temperature of the bath or dry-block.
DATACAL
 
Easy connection system
Easy Connect Push on the terminal top and insert wires with a diameter of up to 3 mm or 10 AWG and compensated thermocouple connectors.

Wires are held tight between two brass plates that provide thermal stability and a very good cold junction compensation for thermocouples.

This system also enables 4 mm banana plugs and security connectors to be connected on the front panel.
  Request For Quote
  PDF THERMYS150 data sheet
  PDF DATACAL data sheet
 
THERMYS150
 
CALYS
Dual display IN / IN or IN / OUT
Screen size: 210 x 110 x 50 mm
Waterproof: IP54
Weight: 900 g only
Operating conditions: 0 / 50° C, 10 / 80% RH
 
Specifications THERMYS150
Channel number 2 channels: IN / IN or IN / OUT
Top accuracy 50 ppm
Thermocouples 0.005% RDG
Resistive probes 0.006% RDG
Thermistors 0.006% RDG
Key features: 50 KOhm range and Steinhart-HART equation
Compliant to standard 21 CFR Part 11 - More information here
21 CFR Part 11, standard from US Food & Drug Administration
Applying to pharmaceutical industry, medical operations and cosmetics, it specifies the way documents and electronic data should be correctly used in order to certify the reliability and safety of a system. The standard was set up to minimize risks related to electronic processing of data.

Key-rules are precisely defined in terms of system validation and safety, audit trail, operational control, documentation management, electronic signature and education.
 
THERMYS150 compliance:
  • Audit trails trace all actions and events during a qualification
  • Access rights are managed by the Administrator
  • Access is authorized through password and user ID given by the Administrator
  • All records are protected in a native format and shall not be corrupted
  • Electronic signature is required for every action into the system
  • Training is provided by the manufacturer

NADCAP Heat treatment
Additional functions Advanced data exploitation: Scaling, relative measurement, simulation of ramps and steps, synthetizer, square root, statistical functions
Drives baths and dry-blocks
DATACAL calibration software
Transmitter function, switch test
Memory up to 10,000 values
Communication USB connexion
Power supply Rechargeable NiMh batteries
 
DATACAL Calibration management software 21 CFR Part 11
100% paperless calibration procedures! DATACAL
Full configuration of instruments
Real time processing of data
Issue your own customized calibration certificates
  DATACAL
 
 
Home |Company | Events | Support |Contact  | © All rights reserved 2008
AOIP SAS - ZAC de l'Orme Pomponne - 50-52, avenue Paul Langevin - 91133 Ris Orangis Cedex - France