THERMYS 150 is a all-in-one thermometer / calibrator for thermocouples, resistive probes and thermistors providing a high accuracy:
- Thermocouples: Up to 0.005 % RDG
- RTDs: Up to 0.006 % RDG
- Thermistors: Up to 0.006 % RDG
Two input channels are available on the instrument for comparison calibration and temperature differential measurement.. Temperature can also be simulated by the output channel in order to control temperature recorders, indicators, handheld thermometers... |
| |
| Advanced features |
- Memory: Store your calibration procedures in the device and save your results for on site easy and quick calibration during the whole week. Then, upload the data on a computer in order to issue calibration certificates
- Customized calibration certificates: Issue your own customized calibration certificates with DATACAL calibration software
(see below)
- Additional functions: Scaling, steps, synthetizer, statistical functions, switch tests... make advanced data exploitation easier
- Driving dry-blocks and baths: Furnaces can be driven by THERMYS150 through a 4-20 mA signal sent by the output channel in order to set up the temperature of the bath or dry-block.
|
 |
| |
| Easy connection system |
 |
Push on the terminal top and insert wires with a diameter of up to 3 mm or 10 AWG and compensated thermocouple connectors.
Wires are held tight between two brass plates that provide thermal stability and a very good cold junction compensation for thermocouples.
This system also enables 4 mm banana plugs and security connectors to be connected on the front panel. |
|
|
|
|
| |
 |
| |
 |
| Dual display IN / IN or IN / OUT |
| Screen size: |
210 x 110 x 50 mm |
| Waterproof: |
IP54 |
| Weight: |
900 g only |
| Operating conditions: |
0 / 50° C, 10 / 80% RH |
|
| |
| Specifications |
THERMYS150 |
| Channel number |
2 channels: IN / IN or IN / OUT |
| Top accuracy |
50 ppm |
| Thermocouples |
0.005% RDG |
| Resistive probes |
0.006% RDG |
| Thermistors |
0.006% RDG
Key features: 50 KOhm range and Steinhart-HART equation |
|
| Compliant to standard |
21 CFR Part 11 - More information here
| 21 CFR Part 11, standard from US Food & Drug Administration |
Applying to pharmaceutical industry, medical operations and cosmetics, it specifies the way documents and electronic data should be correctly used in order to certify the reliability and safety of a system. The standard was set up to minimize risks related to electronic processing of data.
Key-rules are precisely defined in terms of system validation and safety, audit trail, operational control, documentation management, electronic signature and education. |
| |
| THERMYS150 compliance: |
- Audit trails trace all actions and events during a qualification
- Access rights are managed by the Administrator
- Access is authorized through password and user ID given by the Administrator
- All records are protected in a native format and shall not be corrupted
- Electronic signature is required for every action into the system
- Training is provided by the manufacturer
|
|
| NADCAP Heat treatment |
|
| Additional functions |
Advanced data exploitation: Scaling, relative measurement, simulation of ramps and steps, synthetizer, square root, statistical functions |
Drives baths and dry-blocks DATACAL calibration software
Transmitter function, switch test
Memory up to 10,000 values |
|
| Communication |
USB connexion |
| Power supply |
Rechargeable NiMh batteries |
|
| |
| DATACAL Calibration management software |
21 CFR Part 11 |
| 100% paperless calibration procedures! |
 |
| Full configuration of instruments |
| Real time processing of data |
| Issue your own customized calibration certificates |
|
|
 |
|
| |